BRIEF INFORMATION ON ELECTRICAL CIRCUIT OF ELECTRICAL CAPACITY 
SWITCHING UNIT (6M-26)

Capacity switching unit 6M-26 is composed of the following 
components:

a. 	Inside socket (Shch3), specialized for connecting capacity 
switching unit with source of electric current.
b. 	Socket of converting current Shch2, Shch6 specializing in 
dividing feeding network to necessary connection with heaters 
and heating network of apparatus.
c. 	Internal opening socket Shch1 and is used to connect unit with 
apparatus.
d. 	Power channel (adapter) which includes direction equipment 
unit (Y2) and transistor keys (T4, T6) specializing in 
sending necessary power to heating equipment in apparatus 
after receiving signals in apparatus.
e. 	Adjusted power channel which includes condenser C2 and 
entrance filter unit Y1 with transistor unit T3, T1 carried 
on unit's frame, also adapter and unit balancer Y2 with 
transistor unit T5, T2 carried on unit's frame and condenser 
C3 of filter.
		Transistors T7, T6, T5, T3 available in drum CKY 
(diagram 5.123.012) and (diagram 11) with transistor 
available on unit's frame (T4, T6) included in composition of 
apparatus heat adjustment system and unit receiving directed 
signals it will be in a closed status.
		When direction signal enters one of transistors sockets, 
for example T3 (collector [illegible]) the transistor opens 
T6 in capacity switching unit, and heat sensor of apparatus 
is connected to current source. Heaters are connected to 
electrical power when direction signal enters transistor 
socket (in drum).
		Adapter and drum voltage [illegible[ includes each of 
the following: measurement equipment connected to transistor 
T3, primary element [illegible], conversion circuit composed 
of transistor T2 and charging [illegible] C1, key connected 
to a group of equipment on transistor T4, T1 (un drum) and T2 
(in capacity switching unit).

	176
	RESTRICTED

	RESTRICTED

f. 	Filter ([illegible]) which is composed from adapter Tp2 (in 
drum), board CKY, and condenser C1 (in capacity switching 
unit).
		In [illegible] adapter, attached oscillation [illegible] 
is used which is controlled by a key and with help from 
adapter circuit (T2, C2) which forms directed oscillation in 
standard times.
		In the initial stage (project stage) transistors are 
closed, and when power goes in transistor T2 opens and is 
connected to charging circuit of condenser C1. When level of 
power in condenser C1 reaches a level enough to open 
transistor T2, the key opens; the condenser C1 begins 
emptying through transistor T8. At negative power of 
condenser C1 appears on transistor T2 layer and the former 
stays closed until complete emptiness of condenser C1.
		As [illegible] of entering voltage of [illegible] 
adapter from exchange R13-R15, sent to transistor T3 layer 
the main voltage source is considered [illegible], evacuation 
time of condenser C1 is like its charging time related to the 
last calculation of amount of entering voltage.
		By increasing entering voltage time of charging 
condenser C1 is reduced which leads to reducing time of 
transistor T2 remaining closed, as a result voltage leaving 
adapter [illegible] remains fixed because of the reduction of 
oscillation coming from key.
		When reducing incoming voltage the voltage on adapter 
[illegible] exit maintains the operational level. As a result 
of increase in oscillation coming from key and when key is 
open the current enters through filter ([illegible]) to 
resistance whose condenser is charged.
		When key is closed the accumulated energy inside filter 
is used for resistance through transistor T5 (capacity 
switching unit), when valve opens in [illegible] adapter 
circuit it is possible to adjust voltage by changing 
resistance (R14).
		Drum includes adapter (current adjuster) connected with 
transistor T1, T2 in apparatus drum and transistor T1, T3 (in 
capacity switching unit 6m-26) and adapter Tp1 of dual rolls 
which sends

	177
	RESTRICTED

	RESTRICTED

	voltage to current adjuster and filter ([illegible]).
		In order to reduce oscillation influence on regular 
integration circuit +6.5 volts, comparative adapter is used 
which is connected on transistors T7...T3.

CHARACTERS OF CAPACITY SWITCHING UNIT (A 6M.26) (6M-26A)

The difference of capacity switching unit 26 volts (6M-26A) from 
capacity switching unit 26 volts (6M-26) is in necessity of 
sending changing voltage (220/127 volt) to entrance socket Shch3 
and the necessity of connecting it with adapter 220/127 volt as a 
result other keepers are used for safety wire Mp1 and Mp1 and also 
removal of socket cover Shch6 (dividing feeding network to channels 
and resistance and also securing connection [illegible] of heaters 
and heat sensors of apparatus executed by wire connecting unit 
with adapter; characters of connecting and transferring voltage of 
capacity switching unit 26 volt A is similar to an extent 
characters of capacity switching unit 220/127 volt.

DISTINGUISHED CHARACTERS FOR CAPACITY SWITCHING UNIT 26 VOLT B 
(617-265)
Capacity switching unit 26 volt B like capacity switching unit 26 
volt A is different from capacity switching unit 26 volt also in 
the following manner:

a. Wire safety protection (Mp2, Mp1).
b. Absence of socket Shch6 cover.
c. Characters of connecting and transferring voltage.

	178
	RESTRICTED

	RESTRICTED

	(Diagram number 18)
	A general view of capacity switching unit 220/127 volt

[keys]
1) Attachment (adapter) 220/127 volt
2) Main capacity switching unit
3) Division
4) Connection wire























GENERAL INFORMATION ON GENERAL VIEW OF CAPACITY SWITCHING UNTIL 26 
VOLT
This unit was designed in a frame with square angles from 
aluminum, from above and below closed by covers (5, 4) which 
secures protection for the inside from humidity and dust.

On the rear wall of frame there are divisions (7) used to place 
apparatus in operation on the required position and also to 
provide mechanical connection (with help from screws) to unit with 
supports.
The front (2) of capacity switching unit is located into the 
inside with a certain depth and is formed of side surfaces ends, 
and also covers ends where carrying handles (6,3) are located.

On the front, there is:
Entrance socket (receiving voltage) W3
Transferring sockets (Shch6, Shch2)
External keys socket (Shch1).

	179
	RESTRICTED

	RESTRICTED

		Safety wire Mp2, Mp1
		Illustrating light [illegible]
		Place of putting ground wire

9. CAPACITY SWITCHING UNIT 220/127 VOLT (6M 127/220)

Purpose, components and characters
This unit is used to secure the necessary volts when operating 
apparatus from changing current network its oscillation from 50 ± 
0.5 hertz and voltage 220/127 ± 22/13 volts, unit is formed from 
(note diagram 18) capacity switching unit (26 volt A).
Capacity switching unit performs the following functions:

a. Sequential connection for heating and transferring elements for 
apparatus.
b. Connecting sequentially of two of heat sensors.
c. Checking operation of key operation of apparatus (by sensory 
lights).
d. Characters available in table number 8



Table number 8

	180
	RESTRICTED

	RESTRICTED

RESISTANCE 
CURRENT 
(AMPERE)
NAMES 
OF 
CONSUME
D PARTS
ELEMENTS OF CAPACITY SWITCHING UNIT






FREQUENCY
OSCILLATIO
N
VOLTS 
ALLOWE
D 
(VOLT)
REGULA
R 
POWER, 
VOLT



HIGH 
FREQUENCY
; 
ADAPTER; 
VOLT / 
CAPACITY; 
LESS THAN
LOW 
FREQUENCY; 
VOLT / 
CAPACITY; 
LESS THAN



1
2
3
4
5
6

CHANNEL SWITCH 
OF MOBILE MASS 
SIGNALS
0.2-
0.42
-
-
± 6.5
+ 30

CHANNEL OF 
FEEDING WARMERS
2.2-
5.15
-
-
± 6.0
+ 26

CHANNEL OF 
FEEDING HEAT 
POSTS
1.5-
2.70
-
-
± 6.0
+ 26

CHANNEL OF 
FEEDING AUDIO 
SIGNALS
0.006-
0.15
-
-
± 2.5
~ 7.5

CHANNEL OF 
FEEDING 
INTERNAL 
CONTROL BOARD
0.12-
0.18
5
0.9
± 1.0
+ 13

SO
SO
SO
SO
SO
- 13

SO
0.30-
0.40
5
0.1
± 0.2
+ 5.6

CHANNEL OF 
FEEDING MASS 
ELECTRON-LIGHT 
LIGHTING LAMP
1.3-2.0
-
-
± 0.4
~ 4.5

CHANNEL OF 
FEEDING 
APPARATUS DRUM
0.35-
0.45
-
-
± 2.5
+ 28

 
Summary information on electrical circuit of capacity switching 
unit BP- 127/220

This unit is different from capacity switching unit BP-26 by the 
availability of a supplement (adapter BP-127/220), which is an 
adapter from varying current voltage to constant current voltage 
required to operate base of capacity switching unit of apparatus A 
26 volt (BP-26).
The supplemental adapter BP-127/220 is made of the following 
parts, diagram (3.211.005):

	181
	RESTRICTED

	RESTRICTED

-	The Shch1 cavity is used to connect the adapter with capacity 
switching unit (BP-26).
-	The Shch Tr2 regulator is used as an adapter to reduce the 
network's voltage to a low voltage.
-	Voltage switch installed on cavities Shch2 ...Shch4 which are 
used for sequential connection (when the voltage is 220 volt) or 
for parallel connection (when the voltage is [illegible] volt) 
to the primary screw of the [illegible] regulator.
	Current regulator is connected to the transistor T1...T4 
according to electricity connection regulator.
-	The filter is constructed of regulator adapter Tp1 and 
condensers C1...C4.
	Connection of the adapter and supplement PR-127/220 with cavity 
Shch2 of capacity switching unit BP-26 is carried out with 
assistance of wire (6.644.135), cavity Shch1 which secures 
dividing feeding network to channels and resistors, also 
converting connectors of apparatus's heating equipment and 
heating sensors.


A GENERAL VIEW OF CAPACITY SWITCHING UNIT' STRUCTURE (BP-127/220)
	




































[Key]:
1) Capacity switching unit 26 volt, 1
2) (supplement) Adapter-12 volt, 2


A GENERAL VIEW OF CAPACITY SWITCHING UNIT 12 VOLT, (DIAGRAM-19)

	182
	RESTRICTED

	RESTRICTED

		The capacity switching unit BP-127/220 is composed of 
the main unit 26 volt, 1, supplemental adapter PR-
127/220, 2, and are mechanically connected by screws or 
electronically through wire, 4.

		The supplement PR-127/220 is a structure different from 
the capacity switching unit (BP-26) in the front only 
where cavity Shch1 is fixed and in the ground wire 
connection point and voltage switch cover.
		At the front wall there are filter elements, whereas at 
the back wall there is the regulator and drum with 
elements for strengthening the current.

10.	Capacity switching unit 12 volt BP-12

	Purpose, composition, general characteristics

	The unit is used to secure the necessary volts for the 
apparatus when the apparatus is switched from a power source 
with a constant current.

	The capacity switching unit is composed of (diagram-19) 
supplement, adapter, 12 volt and capacity switching unit.

	The unit perform the following functions:
	1.	Sequential connection for heating elements in the 
apparatus.

	2.	Parallel connection for two of the heat sensors of the 
apparatus.

	3.	Displaying the operation light of apparatus switching 
key (with help from the green light characteristics 
illustrated in the table (number 9).













	(183)
	RESTRICTED

	RESTRICTED

TABLE 9
NAMES OF 
CONSUMED PARTS
RESISTANC
E CURRENT 
(AMPERE)
ELEMENTS OF CAPACITY SWITCHING UNIT






FREQUENCY

VOLTS 
ALLOWE
D 
(VOLT)
REGULA
R 
POWER 



FREQUENCY
; 
ADOPTER; 
VOLT / 
CAPACITY; 
(LESS 
THAN)
FREQUENCY
; VOLT / 
CAPACITY; 
(LESS 
THAN)



1. CHANNEL OF 
FEEDING 
SIGNALS
0.2-0.48
-
-
+ 9.0 
- 7.0
+ 26

2. CHANNEL OF 
FEEDING 
HEATERS
3.0-9.75
-
-
+ 3.0
- 4.0
+ 12

3. CHANNEL OF 
FEEDING HEAT 
POSTS
2.5-5.3
-
-
+ 3.0
- 4.0
+ 12

4. CHANNEL OF 
FEEDING AUDIO 
SIGNALS
0.05-0.5
-
-
± 2.0
~ 7.5

5. CHANNEL OF 
FEEDING 
INTERNAL 
CONTROL BOARD
0.12-0.18
5
0.9
± 1.0
+ 13

6. SO
0.12-0.18
5
0.1
± 1.0
- 13

7. SO
0.30-0.45
5
0.1
± 0.2
+ 5.6

8. CHANNEL OF 
FEEDING MASS 
ELECTRON-LIGHT 
LIGHTING LAMP
1.3-2.0
-
-
± 0.4
+ 28

8. CHANNEL OF 
FEEDING 
APPARATUS DRUM
0.35-0.45
-
-
± 
+28

	
Summary information on electrical circuit of capacity switching 
unit BP-12

The capacity switching unit (12 volt) is different from the 
capacity switching unit (26 volt) because of supplement-adapter 12 
volt which converts low voltage of fixed current to voltage equal 
to 26 volts of fixed current necessary for the operating base of 
capacity switching unit 26 volt.

The supplemental adapter 12 volt is made of the following parts:
Cavity Shch used to connect the adapter with capacity switching 
unit 26 volt.
Volume entrance filter (C1, C2) used for dual charging to 
supplement and source of fire,

	184
	RESTRICTED

	RESTRICTED

Voltage adapter includes adapter board PR and (mass Y) and 
transistor T1, T2 carried on structure of capacity switching unit.
The current connected with the transistor T1...T2 according to 
electrical connection regulator.
Volume filter is condenser C3.

The principal of current operation is similar to the operation of 
current adapter 26 volt (paragraph 9). Adapter 12 is connected to 
cavity Shch2 of capacity switching unit 26 volt by wire 
(6.644.136) cavity which secures dividing feeding network to 
channels and resistor, and also heating wires and heating sensors.

A GENERAL VIEW ON THE STRUCTURE OF CAPACITY SWITCHING UNIT 12 VOLT 
BP-12

Capacity switching unit 12 volt (diagram 19) is made of main 
capacity switching unit 26 volt, 1, and the supplement-adapter 12 
volt and are strongly connected. In general it does not differ 
from capacity switching unit supplement-adapter 12 volt installed 
in a unified structure, on the front exists cavity Shch2 and 
points of connecting ground wire.

11. GENERAL VIEW OF EXAMINATION TOOL SET 

Symbol of examination tool set (KMC-E)
The equipment is made of two boxes and used for light chemical 
reaction; each box secures performance of three charges to the 
equipment, each charge has:
A can with examination tape, one piece
Solution bottle number 1, one piece
Solution bottle number 2, one piece
Powder bottle number 1, one piece
Powder bottle number 2, one piece
Powder bottle number 3, one piece
Bottle for categorization solution, one piece
Filter, one piece
Box, one piece

	185
	RESTRICTED

	RESTRICTED

The charges are placed in [illegible] which is put in a wooden box 
with the instructions. Tapes are placed in a metal can.
Time of preparing the sensory solution number 1 and number 2 and 
filling up the dripping tube is no longer than 10 minutes.

Sensory solutions are prepared in temperatures less than 10¡C 
after heating up bottles containing sensory solutions number 1 and 
number 2 inside holes available in the equipment (check diagram 2) 
(24) to temperature below 50¡C.

12. RESERVE MATERIAL SET

The set contains the following tools and equipment used when 
operating the apparatus and for maintenance:

Coils 7.810.0205 screwdriver which is fixed on the apparatus and 
used to change dripping tubes and to open solution bottles and 
powders, screwdriver 8.892.017 is used to change fixed volumes,

Pincers (PA) 150x2.5 and a brush number 30 used when categorizing 
electron-light adapter,

Light filter and board 8.634.266 used to examine and arrange 
measurement equipment 6.062.001 used to examine dripping tube, a 
cotton swap 200x200 used to clean electron-light information and 
also to clean the milky-colored light filter, dose measurement 
8.123.026 is used to measure amount of dose or volume of drip,

Container 8.600.734 used when measuring the dose,

Wire 1-03 its length = 5.1 cm. (10 pieces) used to clean dripping 
tube, rubber tube 2x3 used to change air passage of intake air 
system of the apparatus, light (CG-24-1,2) used to compensate 
inoperative sensory lights of the apparatus and capacity switching 
unit, tick (D[illegible] 247.004) used to compensate damaged ticks 
as a result of friction,

Key 2811002 17x17 used to connect entrance tube with the 
apparatus,

	186
	RESTRICTED

	RESTRICTED

Safety fuse VP-36-1 used to replace damaged fuse in capacity 
switching unit, the following are its different kinds:

Capacity switching unit, BP-127/220
Capacity switching unit, BP-26
Capacity switching unit, BP-2
Light 6.3-15 used to replace damaged light of internal control 
board,
Sensory tape used for purposes of examination,
Filter used to clean air,
Key 8.822.016 used to replace apparatus drum light when damaged.
Note: All the tools mentioned above (except screwdriver 7.810.030) 
are available in reserve material set.
Method of packing and numbering are stated on inside of set's 
cover.

13. MARKS, STAMPS AND PACKING

The primary marks making the apparatus are stated on a specific 
board and on control board and light signals of the apparatus, and 
on front of mobile signals volume and on front of mobile signals 
volume and on front of capacity switching unit.

On boards fixed on the apparatus and on capacity switching unit 
are the following information:

Mark of apparatus manufacturer,
Manual (1 cm.) of apparatus,
Factory serial number of apparatus,
Year of manufacture
On mobile signals, the volume board is the mark of the manufacture 
of the apparatus with the manual of apparatus. In addition, on the 
capacity switching unit, the board is the unit's manual under the 
apparatus manual.
Mark on box is written with black paint that cannot be erased.

	187
	RESTRICTED

	RESTRICTED

Also on the side front of the box are the following symbols, "an 
upward arrow, a cup and umbrella." Factory code is also written, 
laden form, production number, address of factory, number of 
inventory, stamp number and weight; in addition to that, there is 
on the diagonal section of the box the marks and symbols "an 
upward arrow, a cup and umbrella."

Inside packing box are all parts of the apparatus including the 
apparatus and capacity switching unit. The box is sealed. To seal 
the box, there are designated holes on apparatus cover key through 
which the seal is connected.
Sealing capacity switching unit is done by special material which 
fills a designated hole placed at place of connecting the upper 
and lower covers with structure (box) of apparatus.

Packing box is used to maintain the apparatus when moving and from 
long term storage.

Placed inside packing box are:
Apparatus, mobile signals volume, capacity switching unit, reserve 
material set, examination tool set, tubes for entrance and exit, 
usage manual, instruction manual, service record and packing 
paper.

To protect the apparatus from humidity, all parts of the apparatus 
will be packed in carton boxes and plastic bags (polythene) 0.2 
ml. thick before packing in the box, and to absorb air humidity, 
cloth bags containing [illegible] powder will be placed.

NOTE:	Packing box will be returned to the consumer after depositing 
the apparatus with him.

14. SAFETY PRECAUTIONS

Used sensory tape and anti [illegible] filter will be replaced, 
and humidity container will be cleaned using protective gloves 
taking into consideration safety precautions of nerve agents 
vapors if used in a contaminated ground.

Care must be taking to prevent dripping of cleaning liquid and 
solution number 2 containing methanol (poisonous) on [illegible] 
glands and breathing passage, if these parts have been 
contaminated they must be washed immediately with cold water.

After finishing work with stated liquids, hands must be washed 
with water and soap.

The apparatus will be turned on only after verifying the 
connection of ground wire when using, an alternating current 
circuit with a voltage about 127/220 volt.

	188
	RESTRICTED

	RESTRICTED

Take precaution against high voltage electric shocks if the 
apparatus was working on 127/220 volt with alternating current. If 
it was necessary [illegible] with the apparatus and opening the 
front board of the apparatus or covering capacity switching unit.

15. METHODS OF PLACING THE APPARATUS

a. General instructions::

Remove covers of entrance and exit holes of the apparatus and 
covers of air passages and place these parts in the reserve 
material set. Place the apparatus on the necessary supports 
available at the designated place, and capacity switching unit 
will be placed at the designated area. Place special cans of 
sensory tools KIS and reserve material set inside the designated 
area around the apparatus.

Using wire 6.604.134 connect cavity 12 in the apparatus (diagram 
1) with the free cavity Shch1 in capacity switching unit. Place 
mobile signals volume (PVS) in the designated area (if it was 
available with the apparatus) and connect them using wire 
6.664.13[illegible] or 6.664.13[illegible] with cavity of 
apparatus.

b. Apparatus is installed at the designated area with "by bass" 
system or linear system or traffic system (form [illegible] and 
form 2).
Connect entrance and exit holes of apparatus with entrance tube 
6.412.016 and exit tube 6457.050.

Keep cavity cover 15 away from apparatus (diagram 1) and connect 
with it wire belonging to entrance tube.

Connect entrance and exits tubes with appropriate parts of traffic 
system [illegible] to secure good connection.

Turn off the main operating switch 2 (diagram 1) of the apparatus; 
connect the main network wire of the location with cavity Shch3 
available in capacity switch unit 26 volt (BP-26).

	189
	RESTRICTED

	RESTRICTED

c. Install apparatus in designated area without system "by bass" 
(V-form).
Connect main network wire with cavity Shch3 available at capacity 
switching unit 12 volt.

NOTE: 
When installing the apparatus at the designated area equipped 
according with B-form, it is necessary that it be connected with 
network 170/220 volt through wire 127/220 (6644 37) and ground 
wire.
Wire 127/220 is connected with cavity Shch3 of apparatus of 
capacity switching unit BP-127/220.

d. Points with implementation dependent on temperature of the 
surrounding air if air temperature was over 20¡C, heating switch 
must be turned off.

NOTE: If temperature of surrounding air was between 10-20¡C, it is 
better to use the apparatus without turning on heating switch.
If temperature of surrounding air is less than 10¡C, heating 
switch should be turned on.

16. PREPARING APPARATUS FOR OPERATION

NOTE: Care should be taking to make sure that there are no covers 
in air intake vent and vents of heat sensors fan available in the 
back area of the apparatus.

GENERAL INSTRUCTIONS

Before anything and when preparing apparatus for operation make 
sure that the apparatus is insulated from heat and appropriate for 
proper operation; if the insulation of apparatus is not suitable 
for proper operation, apparatus must be prepared properly and 
adequately for proper operation by following the following 
necessary instructions:

	190
	RESTRICTED

	RESTRICTED

Preparing apparatus for operation includes several points in the 
following order:

a.	Operating the apparatus
b.	Placing solution bottles in the apparatus (when temperature 
of surrounding air is less than 10¡C).
c.	Placing the sensory tape.
d.	Placing filter in the apparatus.
e.	Heating the apparatus to the required temperature.
f. 	Washing dripping tubes.
g.	Filling up dripping tubes.
h.	Checking operation of system of pulling tape and dripping 
tube.
i.	Final heating of the apparatus.

		NOTE: When temperature of surrounding is more than 10¡C 
(and in order to shorten time required for preparing 
apparatus for operation), points "f", "g" and "h" should 
be executed after point "d".

OPERATING APPARATUS

Switch operation key of apparatus 5 (diagram 1) at operation 
position BKL.

PUTTING SOLUTION BOTTLES IN APPARATUS

Solution bottles are put in apparatus for the purpose of heating 
them if air temperature was less than 10¡C, to do that:
Open box of sensory tools set KIS and take out sensory solution 
bottles number 1 and number 2 and also cleaning solution bottle.
Open apparatus 2 cover (diagram 1) by opening keys 10 and 14.
Pull restriction bow 16 (diagram 2) toward [illegible] then turn 
it around in a way to facilitate placing bottles inside the proper 
hole in apparatus, after that return restriction bow to its 
original position.

	191
	RESTRICTED

	RESTRICTED

PLACING SENSORY TAPE

To place sensory tape in the apparatus, the tape must be taken out 
from sensory tools set KIS.

Open tape's container and take out tape's roll and cut about 4-5 
cm.; then place sensory tape on sending dial in a way to secure 
movement of the tape clockwise. Open placement lever of heat 
blocking board of the tape in which the board will be moved 
downward, and pull placement lever 12 to the right. Roll the cut 
tape according to illustration on the inside of apparatus cover, 
then place the loose end of tape on reception dial teeth and turn 
it twice or three time clockwise; then [illegible] tape by turning 
reception dial clockwise and return placement lever to its 
position, and also do the same for placement board after placing 
it using the lever.

PLACING FILTER

To place filter, perform the following:
Take out filter from sensory tools box. Keep placement lever of 
filter 25 (diagram 2) upward and place filter in hole resulting 
from lifting lever upward, then release lever so that it returns 
to its original position.

HEATING APPARATUS TO NECESSARY TEMPERATURE REQUIRED FOR OPERATION

To raise temperature of apparatus to the required temperature, 
perform the following:
Put key 18 (diagram 2) (preparing-operation) on Operation position 
and area selecting key on Cyclical position.
Press key of pulling tape system and close apparatus cover. Let 
the apparatus work until the blue light is on.

NOTE:
If temperature of surrounding air was more than 32¡C, the blue 
light may turn on immediately after operating apparatus when key 5 
is switched. This indicates that apparatus is ready for operation 
without a need for heating.

	192
	RESTRICTED

	RESTRICTED

WASHING DRIPPING TUBES

To wash dripping tubes, perform the following:

Open apparatus cover and open placement lever of heat blocking 
board where it will go downward.

Pull holder 10 of right dripping tube 11 (diagram 2) and turn it 
to the side; then carefully pull dripping tube toward you after 
liberating it from base and holder.

Pull restriction bow 16 (diagram 2) toward you. Take cleaning 
solution bottle and then return restriction bow to its pervious 
position.

Open nut 15 (diagram 9); and open dripping tube cover, turn it to 
the side. 
Pour cleaning solution in dripping tube cup to level of 
[illegible] bow of dripping tube, then close dripping tube cover. 
Shake it five times, then move axle 11 by hand upward and 
downward. When axle is moved downward, a stream of liquid should 
come out from [illegible], verify that.

Lift dripping tube cover and turn it to the side. Empty dripping 
tube of cleaning solution, then move axle 11 upward and downward 
until you are sure that solution has been completely drained. Wash 
right dripping tube again using fresh solution and following the 
same procedure.

Close dripping tube using its cover 6, tie nut 15 and place 
dripping tube in its position. Then place it using holder 10, pull 
holder x of left dripping tube (	) in diagram 2 toward you then 
to the side. Take dripping tube and then wash left dripping tube 
the same way used with right dripping tube, close apparatus cover 
afterward.

FILLING DRIPPING TUBES

To fill dripping tube, perform the following:

Open apparatus cover, pull restriction bow 16 (diagram 2) and take 
bottle containing solution number 2; then return restriction bow 
16 to its pervious position. Prepare sensory solution number 2 
(green color) and follow instructions stated in manual for period 
of sensory tools. Pull right dripping tube according with above 
instructions.

	193
	RESTRICTED

	RESTRICTED

Open dripping tube cover in dripping tube cup where there is 
prepared solution filled up to [illegible] level of dripping tube 
cup, then close dripping tube cup with its plastic cover. Shake 
cup to wet cup's inside. Then move axle upward and downward four 
or five times, and then empty cup of solution. After that move 
axle downward and upward and make sure that cup is empty of all 
solution. Then lift right dripping tube cover and turn it to the 
side and pour what is left of solution in cup. Close dripping tube 
cover and return it to its place.

Prepare sensory solution number 1 (transparent) according to 
instructions and fill up left dripping tube using the same method 
followed above with right dripping tube.

NOTE:
After filling up dripping tubes and when necessary, put new 
bottles in designated holes for heating up solution bottle's in 
place of the ones used.

EXAMINING VALIDITY OF OPERATION OF SYSTEM FOR PULLING TAPE AND 
DRIPPING TUBES

For this purpose, perform the following:

Move sensory tape by pressing key 15 (diagram 1), then make sure 
of movement of sensory tape and operation of dripping tube at the 
same time.

Examine amount of drip coming out of each dripping tube as 
follows: place key (preparation-operation) on Operation position, 
and place selection areas key on Constant position.

Press key Examine (KONTR), and with help from a special sample 
8.600.734 available in reserve material set, collect examiner. 
After that pull sample from under dripping tube, press sensory key 
noting indicator arrow in apparatus. After arrow returns to point 
zero, remove sensory tape. With help from drip size measure 
6.59.00 taken from reserve material set, measure sensory 
evaluation on sensory tape. If darkness dimensions formed by 
sensory solution from each dripping points was less or more than 
dimensions indicated on drip size measure, drip size should be 
changed through the following:

	194
	RESTRICTED

	RESTRICTED

Without taking it out open, screw 7 of restricting ring 9 (diagram 
9) by screwdriver.

Turn screw 8 times from 1-0.25 counter-clockwise if it was 
necessary to reduce size of drip and opposite (clockwise) if it 
was necessary to increase drip size.
Measure drip size again, if size of hole available on sensory tape 
matches the size, screw 7 will place restricting ring 9, then 
place sensory tape. Place switch (preparation-operation) on 
Operation position, left heat blocking board and close apparatus.

NOTE:

1.	When washing and filling up dripping tubes, the blue light is 
allowed to be off.

2.	After measuring drip size, it is allowed to clean dripping 
tube bead by a wire from reserve material set and pincer, 
then measure drip size again.

FINAL HEATING OF APPARATUS:

To perform a final heating for apparatus, perform the following:
Place key 4 (diagram 1) on Cyclical position; from one time to 
another, press on tape-pulling system key and wait until sensory 
lights are on. After that the apparatus would automatically 
operate in the cyclical position.

ATTENTION:

When preparing apparatus for operation, the apparatus may give a 
danger signal and sensory lights go on. Danger signal and 
dysfunction should be given only after apparatus went through 
final heating.

NOTE:

1.	When preparing apparatus for operation in conditions where 
temperature of air is above 10¡C, it is allowed to wash 
dripping tubes and fill them up, check tape and dripping 
tubes pulling system and above operations without heating up 
apparatus to the required temperature.

2.	When preparing apparatus for operation, it is preferred (for 
purposes of speeding up raising temperature inside thermostat 
section) to close apparatus cover while preparing sensory 
solutions and washing, filling up and adjusting dripping 
tubes.

	195
	RESTRICTED

	RESTRICTED

17. OPERATION PRINCIPAL

Attention! When air temperature is less than 10¡C, apparatus 
equipped with sensory tools set should be always open (heating key 
open); and when air temperature is more than 20¡C, heating key 
should be closed.

When discovering components of organic phosphor vapors in 
operation system Cyclical of apparatus, apparatus should be 
changed to constant operation system as soon as possible.

Perform switching from one system to another is also on tape-
pulling system key.

One person operates apparatus who is very well trained. Apparatus 
is designed for work appropriately and constantly without a need 
to refill dripping tubes during:
8 hours when in constant system
24 hours when in cyclical system.

In order to switch apparatus from one system to another (from 
cyclical operation system to constant operation system), it is 
necessary to change position of work area selection key 4 (diagram 
1) to the needed position (constant-[illegible] for constant 
operation, cyclical [illegible] for cyclical operation).

And press on tape-pulling system; after that apparatus will 
operate according to selected system.

Signals indicating existence of organic phosphor components (nerve 
agents) and after switching work areas selection key and until 
pressing on tape-pulling system will not be taken into 
consideration.

It is necessary to note mobile signals mass and light board in 
apparatus front (diagram 2) and especially on control board and 
light signals 7 of apparatus while in operation. Apparatus 
operating properly should have blue light constantly on.

Green light will be off from time to time according to operation 
time of tape-pulling system (while apparatus is working according 
to constant system during each 120 ± 10 seconds, and while 
apparatus is working according to cyclical system during each 120 
± 10 seconds and then 840 ± 30 seconds, after that 120 ± 10 
seconds...etc.).

	196
	RESTRICTED

	RESTRICTED

If blue light available on light board did not go on, it is 
necessary to check it by pressing on key 15 (diagram 1) of tape-
pulling system. If above mentioned key was pressed and light did 
not turn on, light should be replaced. 
When blue light goes on when pressing tape-pulling key it is an 
indication that the automatic adjustment system for heat is 
inoperative. In this case, apparatus should be repaired.

When apparatus detects nerve agents vapors, yellow light available 
on mobile signals mass and on light board will blink (on-off), at 
the same time siren will go on.

If concentration of poisonous vapors continue to be higher than 
apparatus sensitivity of poisonous agents, danger signal will 
repeat every two minutes when apparatus is working according to 
constant system and every 16 minutes according to cyclical system. 
If signal repeating was given in a way that some cycles do not 
give signals ([illegible]), this means that concentration of 
poisonous vapors matches with apparatus sensitivity of these 
concentrations.

If red light began blinking (on-off) when working, this indicates 
that sensory tape is finished or is cut. In this case, apparatus 
cover should be opened. If you verify that the tape is finished, 
place another tape according with above instructions. If tape was 
cut, remove reception dial wheel 15 (diagram 1), take out tape 
being used, pull end of tape from sending dial for 40-45 cm. and 
place tape's loose end on cavity of reception dial wheel and turn 
it 2-3 cycles clockwise. Then place wheel on reception dial axle 
and then turn reception dial until you make sure that tape is 
tight.

When apparatus is in operation, observe red light indicating 
apparatus breaking down which could be prevented following 
instructions in technical maintenance instructions.

When finishing operating apparatus, turn it off and open apparatus 
cover and remove used section of sensory tape and filter and put 
them in the designated place. Pull dripping tubes and humidity 
absorbing container, and pour remaining solutions in designated 
place. Then wash them according to above instructions and place 
them in their place.

	197
	RESTRICTED

	RESTRICTED

Check the apparatus visually and if there is no damage to it, 
close apparatus cover.

The apparatus is cleaned by wiping the external structure with a 
piece of wet cloth wet with purifying solutions number 1 and 
number 2 with organic and containing ammonia.

Preparing apparatus in accordance with external conditions and 
weather:

After opening apparatus cover in short intervals (less than five 
minutes) and apparatus is operating, it is important to press 
tape-pulling system within short times and wait for blue light to 
go on. After that continue operating apparatus.

When air temperature is over 10¡C, it is allowed to operate the 
apparatus discontinuously. But the time necessary to keep sensory 
solutions in dripping tubes should not exceed 24 hours. After 
these periods are over, dripping tubes should be filled with fresh 
solutions.

When turning the apparatus on after a stop from 15 minutes to 2 
hours, it is necessary to press tape-pulling system key and, after 
the light is on, continue operation.

If period of stop exceeded two hours, apparatus should be heated 
and its validity for operation is checked according with above 
instructions.

When teaching students the (teaching period) method for using 
apparatus, it is allowed to use sensory tape available with 
reserve material set instead of original tape and distilled water 
to fill up dripping tubes.

NOTE:	It is not advised to open apparatus cover while in operation 
except in emergency.

Apparatus cover should be closed when apparatus is in use.

18. CHECKING TECHNICAL CONDITION OF APPARATUS

Technical condition of apparatus is checked when delivering 
apparatus to units for practical use and when sending apparatus to 
storage and when maintaining apparatus.

Table number 10 includes the most important points to be 
implemented. Validity of apparatus for operation is determined by 
operating while filled with sensory material. Time necessary to 
measure technical variables (no less than an hour). While 
measuring variables, dripping tubes are filled with distilled 
water and sensory tape is placed and measurement is performed 
while apparatus cover is open.

	198
	RESTRICTED

	RESTRICTED

When checking, visually note the important mechanical sections. 
(such as dripping tubes, tape-pulling system...etc.). If signals 
lights and lighting were inoperative, they are replaced from 
reserve material set and checking will be continued.

Apparatus is technically valid for operation if technical 
variables match descriptions available in table (number 6). After 
checking, wash dripping tubes and clean humidity collection 
container and write down observations in designated sections.

NOTE:	Dripping points adjustment id performed according with 
instructions in table 10.

Table number 10
Most important checks regarding apparatus technical condition.

SECTION CHECKING TOOLS AND 
EQUIPMENT NECESSARY FOR 
CHECKING, CHECKING METHOD
TECHNICAL REQUIREMENTS

	1
	2

1. Check operation sequence and 
their time while apparatus in 
operation by observation and 
[illegible] one element section 
of sensory tape. System 
selection key position 
determines area apparatus 
operates on. Checking is to be 
conducted in both constant and 
cyclical positions and 
(preparation-operation) key 
should be set on Preparation. 
All checks are conducted after 
pressing pulling and turning 
system of tape.
Continuation of air intake is 
determined as follows:

a.	When apparatus is 
operating according to 
constant system, from 
moment of placing tape in 
air intake room until it 
is liberated.

b.	When apparatus is 
operating according to 
cyclical system, from 
moment of operating 
cyclical pump until it is 
stopped.
When apparatus is operating, 
the following points should be 
implemented sequentially 
according to time specified for 
each section of sensory tape 
sections.

a.	Intake air to be analyzed 
through section of 
operating tape by 120 + 10 
seconds.

b.	Sensory tape section is 
moved through 120 seconds 
after beginning of air 
intake.

c.	Wetting sensory tape with 
solution number 1 through 
3 ± 1 second after moving 
tape.

d.	Wetting sensory tape with 
solution 2 through 120 ± 
10 seconds after moving 
tape.

e.	Measuring light intensity 
of sensory tape section.


	199
	RESTRICTED

	RESTRICTED

	1
	2

2. Tape pulling system is to be 
checked as follows:
Open apparatus cover, place 
work area selection key 
(cyclical-constant) on constant 
position and (preparation-
operation) switch on Operation 
position. Turn on apparatus 
switch and press checking key, 
in times between operation of 
tape-pulling system place on 
sensory tape 3-5 drips (dark 
colored spots) on right and 
side section of intake room.

Measure distance between spots, 
between each spot and another 
(pay attention to 
appropriateness of tape 
rolling, organizing tightness 
of tape is done by crew driver 
through turning fixation nut.

3. Checking operation of 
dripping tubes is performed 
with help of transparent glass 
6.059.001 containing bead 
holes.

To measure drip size, put 
transparent glass under 
dripping tube and even in air 
intake room when dripping tubes 
are operating normally, 
solution will pass freely 
through transparent glass holes 
without falling on its surface.
Tape-pulling system should 
secure tape moving a distance 
equivalent to 25 ± 1 ml.

Loosening of sensory tape is 
not allowed.















Dripping tubes should not leak. 
Dripping solution of dripping 
tubes should be in middle of 
tape.
Solution is allowed to drop on 
its surface so that total size 
of dripping solution 2 micro 
litre and should not exceed 
this limit.






	200
	RESTRICTED

	RESTRICTED

	1
	2

Size of drip coming out from 
each dripping tube is measured 
by sample 6.600.734 and level 
measure 123.026. After 
adjusting, leave first work of 
dripping tube and measure drip 
size after second work.
Remove right dripping tube, and 
remove [illegible] filter. 
Color sensory tape by dark 
spot, red light will then turn 
on constantly. Place dripping 
tube and filter in their 
places.

4. Sensory lights will be 
checked when apparatus cover is 
open.
Turn on operation switch 5 
(diagram 
1) and make sure that green 
light on apparatus board and 
mobile signals mass is on.

Press tape-pulling system, blue 
and yellow and red lights will 
then turn on (if they were 
valid). They appear on system 
board and mobile signals mass 
without depending on sensory 
circuit they took place in. 
Note that red light blinks.

5. Checking practical capacity 
of signals circuits regarding 
validity of apparatus is done 
through area selection key
Size of drip coming out of each 
of the two dripping tubes 
should be around 30-40 micro 
millimeter.










Apparatus should provide:
a.	Light signal (through 
green light) indicating 
reception of voltage.
b.	Green, blue and yellow 
sensory lights can be 
checked.









In apparatus there are the 
following light signals that 
indicate problems in apparatus:




	201
	RESTRICTED

	RESTRICTED

	1
	2

To operate on Constant 
position, open network key and 
block opening of air entrance 
22 (diagram 1), red light 
should then be on constantly. 

Remove sensory tape from light 
measurement area, red light 
should then blink regularly.

6. Sensitivity of apparatus 
operation is checked and 
adjusted by sensory light 
filter. Thus remove dripping 
tubes (to prevent wetting tape 
by sensory solutions).

Place area selection key 
(cyclical-constant) on Constant 
position, and (preparation-
operation) key on Operation 
position. 

Place protective screen 
8.634.766 and press key Check 
and adjust measurement 
indicator 22 where it should 
mark an average of 1 ± 6 micro 
ampere so that it suits 
sensitivity of apparatus 
operation (after leaving Check 
key, indicator should return to 
zero). Place sensory light 
filter after 5-10 seconds from 
the moment tape-pulling system 
engine stops (this indicates 
red light turning on) then 
continue observing measurement 
indicator.
a.	Constant red light, air 
intake into apparatus is 
more or less than level 
required.
b.	High heating for tape 
wetting area.
c.	Red light blinking, tape 
is over or is cut.













Apparatus should be adjusted to 
sensitivity of apparatus 
operation in which the 
indicator will move to: 
1 ± 6 micro ampere when 
pressing Check key (actual 
sensitivity of apparatus 
operation); 1 ± 6 micro ampere 
when placing sensory light 
filter in light measurement 
area.
Absolute amount of differing 
indicators by Check key and by 
light filter should not exceed 
1 micro ampere.



	202
	RESTRICTED

	RESTRICTED

	1
	2

Adjust measurement indicator 
and note the arrow, it should 
point to 1 ± 6 micro ampere. 
This is suitable to sensitivity 
of apparatus operation.

If absolute amount of differing 
first and second indicators 
more than 1 micro ampere, 
through sensitivity regulator 
20 (diagram 2) sensitivity of 
apparatus is regulated (turn 
screw clockwise if sensitivity 
was more then needed)

Repeat the operation several 
times until you are sure that 
the absolute amount of 
difference when pressing on 
Check key by using sensory 
light filter is no more than a 
micro ampere.




NOTE:

When checking and adjusting sensitivity of apparatus operation, 
extreme care should be taken and light analysis area should not 
come under direct or reflected sun light.



	203
	RESTRICTED

	RESTRICTED

19. 	POSSIBLE BREAKDOWNS AND FIXING THEM

	Possible breakdowns, their reasons and ways pf preventing 
them are listed in table 11. A breakdown in apparatus is to 
be fixed after 10-15 minutes of turning off apparatus (when 
emptying out condensers). If it was necessary to find out the 
breakdown in apparatus while it is operating, remember that 
apparatus has a high voltage.
	To replace safety wire, open cover (holder) of capacity 
switching unit and take out damaged safety wire and replace 
it with a new wire from reserve material set suiting voltage 
of apparatus.
	For voltage 170/220 volts, safety wire B 35-1-2A (1.25 
ampere)
	For voltage 26 volts, B 35-1-5A (5 ampere)
	For voltage 12 volts, B -36-1-10A (10 ampere)
	Then install cover (holder).
	To replace sensory lights, open light cover (light cover) and 
take out damaged light and replace it with a new light and 
close light cover (on control board and light signals, metal 
protector and vision glass should be removed). To change air 
volume measuring light, remove cover 5 (diagram 2), turn 
screw and remove light cover, remove damaged light and 
replace it with a new light, return light cover and close 
measuring cover.
	To change friction [illegible], turn (by screw driver) 
adjustment nut, remove [illegible] nut and also cover from 
vibrator and replace leather [illegible] with another from 
reserve material set; after that reinstall installation area 
reversing disassembling process and arrange it as indicated 
in maintenance operations.
	Changing rubber tubes (air entrance tubes) is carried on 
inside repair shop.
	To change electron-light adaptor light: open four screws and 
remove cover. Open two screws of light unit and remove 
damaged light cover; using a screw driver, open tow screws on 
side part of cover pressing light enclosure and replace light 
then press enclosure with new light with necessary screws and 
then install cover in light unit.
















	204
	RESTRICTED

	RESTRICTED

	The light should be placed in where channel line drilling 
axle is perpendicular to visual axle of condenser, and the 
same channel line is placed to be similar in terms of visual 
condenser axle. Turn screws of light unit and place cover in 
its place.
	Open apparatus and place sensory tape. Make sure that the 
lighting is equal and is oviform for wetting on sensory tape. 
Changing distance between visual system lenses is done by 
turning lens frame of light unit which is located under light 
unit cover.
	NOTE: Changing electron-light adapter lighting lamp should be 
conducted be a specialist in repair shop conditions.







































	205
	RESTRICTED

	RESTRICTED
@@@
Table 11
DAMAGE, SIGNS AND ADDITIONAL INDICATIONS
1 PROBABLE CAUSE
2 REPAIR
3 NOTES
41. 	Green light does not go on control board and mobile signals 
unit when turning on network switch. Safety wire damage. 
Replace Safety wire.2. One of the green light does not go on 
control board or mobile signals unit when turning on network 
switch.Damage to lamp itself.Change lamp. When opening 
apparatus cover (apparatus front door), front side lighting 
lamp does not turn on.SoSo4. When pressing tape-pulling key 
system, one (or more) sensory light does not go on: blue, 
yellow, red on control board or mobile signals unit.SoSo5. 
Red light is constantly on and air amount measurement cork is 
located at bottom of measurement.Amount of air in air 
passages is less than adequate, [illegible] filter is dirty, 
pump is dirt or broke down.Change [illegible] filter, pump 
breakdown is fixe in repair shop.6. Red light is constantly 
on, air amount measurement channel is not lit.Measurement 
light is burned.Change light 7. Red light is blinking. 
Sensory tape is over.Put a new sensory tape in apparatus.8. 
Red light is constantly on, air amount measurement cork is 
located at top of measurement.Automatic adjustment system of 
air is broken down.Repair at repair shop
	206
	RESTRICTED

	RESTRICTED

12349. Red light is constantly on, air amount measurement cork is 
located at middle of measurement. Sensory tape light in light 
analysis is not available.Electro-light adapter lamp is not 
working. Change light 10. Red light is constantly blinking, air 
amount measurement cork is located in middle of measurement. 
Sensory tape light in light analysis area is available.High 
[illegible] of sensory tape wetting area.Repair in repair shop.11. 
Differences in operations sequence and their execution in 
apparatus.Breakdown in programming system.Repair in work shop 12. 
Heaters and heat sensors are not operational.Breakdown in 
automatic heat adjustment system.So13. Fan does not work.Fan is 
broken or heat regulator circuits are broken. Damage to placement 
nut.So14. Air heaters are not working.Heating coil is 
broken.Change nut.15. Tape vibration is not possible. Nut is 
damaged. Repair in repair shop.16. Tape-pulling system is 
operating independently without stopping.[illegible] key is 
broken.Repair in repair shop.

	207
	RESTRICTED

	RESTRICTED

20. TECHNICAL MAINTENANCE AND ADJUSTMENT
Apparatus is maintained to keep it in a good operational condition 
and should be conducted by qualified individuals.
Types and interims of maintenance are available in table 12.
Signal "+" in table indicates type of maintenance executed.
Signal "-" indicates work (maintenance) which is not executed.
Table number 13 contains all works necessary for all types of 
maintenance; it also contains the technical condition apparatus 
components should be in; and also the tools used for maintenance 
and checking method.
In maintenance and when necessary, repair of apparatus should be 
conducted to prevent all breakdowns and damages available in 
apparatus.

Table 12
1) Type of operation
2) Maintenance
3) Time necessary to maintain one apparatus
4) Apparatus available for use or are stored in storehouses
5) Apparatus stored for long term
6) Daily maintenance
7) Type
8) one [illegible] a year

1. Checking outside appearance of apparatus+++15 minutes 2. Making 
sure that all components of apparatus are available.+++15 minutes 
3. Checking technical condition of apparatus-++1.5 hour 4. 
Registering working hours in service registrar.+++10 minutes 5. 
Replacing expired components -+-20 minutes 6. Washing gas 
passages-++30 minutes 7. Storing and greasing apparatus-++20 
minutes


	208
	RESTRICTED


	RESTRICTED

If damage was unavoidable, the apparatus is to be repaired in 
repair shop at the beginning of each year; and times of operation 
of last year should written in service registrar.
Replaceable components after their expiration are written in table 
2.

Table 13

Works performed to all types of maintenance
WORK CONTENTS AND EXECUTION METHOD TECHNICAL REQUIREMENT 
EQUIPMENT, TOOLS AND SUPPLEMENTS NECESSARY TO PERFORM MAINTENANCE 
OPERATIONS 1231. General view of apparatus that contain:
a. Checking condition of placing apparatus and tubes, entrance 
tube, exit tube, wires and cavity.
b. Check paint condition of outside surface and condition of 
sensory lights cover.
c. Open front cover and make sure screw driver, dripping tubes and 
other material are available.
d. Make sure there is no damage in placement hole of apparatus 
cover.
e. Remove humidity, dust and dirt by brush from reserve material 
set; rust should be removed by a rag wet with gas.External 
components of apparatus should be clean from dirt, rust, scratches 
and other types of damages.Brush; rag; gas

	209
	RESTRICTED

	RESTRICTED

1232. Making sure of apparatus components is done by special 
registrar of apparatus. Attention should also be paid to reserve 
material set and their packaging method. Check sensory tools set 
and check their manufacturing date and their contents according to 
instructions. Also check glass containers and their condition. 
Also check condition of powders.
After checking contents of set, they should be packaged well to 
prevent any damage when moving; complete missing tools and 
equipments. As for sensory tools set, they should be replaced when 
they expire.Components of apparatus must be complete and without 
shortage according to instructions; also sensory tools set 3. 
Check technical condition of apparatus in accordance with 
instructions.Apparatus should be operational.Timing clock; 
automatic pen; light filter; examiner; distilled water; sensory 
tape 4. Daily operation time of apparatus is registered in service 
book of apparatus; when a year have passed, the total time of 
operation should be counted.Daily operation time of apparatus and 
yearly operation time, total daily operational time within a year 
should be registered in time and accurately.5. Material and tools 
that expired should be replaced with new ones from reserve 
material set.
(A table of material and method of replacement is available in 
table number 2).

	210 
	RESTRICTED

	RESTRICTED

1236. Washing gas passages in done as follows:
a. Take anti [illegible] filter from reserve material set and 
place it in apparatus without placing sensory tape.
b. Take from reserve material set rubber tube 2x3, insert one end 
in container filled with alcohol and place the second end in air 
entrance hole of apparatus. In exit hole place container to 
treated alcohol.
c. Turn on apparatus and withdraw through air passages about 
100-150 millimeter of alcohol. After that remove rubber tube from 
air entrance of apparatus.
d. After washing gas passageway, apparatus should be turned on for 
two hours and on constant position to get rid of all remaining 
alcohol.Alcohol; [illegible] filter; rubber tube. (from reserve 
material set).7. Lubricating apparatus should be made according to 
instructions.All non-painted components should be oiled; also 
apparatus should be placed inside nylon bags inside them 
[illegible] bags.Anti rust oil; rag and gas.


	211
	RESTRICTED

	RESTRICTED

21. METHOD OF TRANSPORTING, STORING AND LUBRICATING APPARATUS
Apparatus is transported inside a special box from manufacturer by 
all types of transportation and at any time. Shipping and handling 
is done in accordance with instructions written on box. 
Transportation is done in closed transportation. If transportation 
was open, a special [illegible] is used for protection; when 
apparatus is transported, box is firmly placed to prevent movement 
of apparatus and prevent shocks. Apparatus is stored in closed 
warm areas with temperatures between 5-40 degrees celsius and 
relative humidity is less than %80.

NOTE: It is allowed to store apparatus (without sensory tools set) 
in unwarmed areas and with weather temperature less than 40 
degrees celsius and relative humidity less than %95.
Sensory tools set are stored according to signals written in 
manual (illegible). When storing for prolong periods, apparatus 
should be checked regularly at least once a year and check 
apparatus lubricating and when necessary (cracks appearing in 
nylon bags, disappearance of [illegible] color, appearance of 
erosion). Apparatus is to be relubricated after transporting or 
storing apparatus; in cases of below zero temperature degrees, 
apparatus should not be operated. In cases of above zero 
temperatures, it should not be operated until after two hours in 
those conditions.
Apparatus is stored in work area for three months as follows:
a. Technical maintenance of apparatus in accordance with 
instructions in table 12.
b. Place clean piece of [illegible] from sensory tape between 
moving and fixed part of air intake room.
c. Close cover and lubricate all non-painted outside surfaces with 
a light coat of rust resistance oil.
d. Write down executed procedure in service book.

Stored apparatus is removed as follows:
a. Remove erosion resistance oil by rags then maintain apparatus 
according to table 13.
b. Write down important notes in service book.
Store and maintain apparatus stored for long term is done after 
one year of storage or after wetting sensory bag (blue color 
changing to pink according to the following points:

	212
	RESTRICTED

	RESTRICTED

a. Open package by cutting top surface.
b. remove apparatus from carton package.
c. Remove erosion resistance oil by rags.
d. Change wet [illegible] with a dry one.
e. Place with apparatus and capacity switching unit an absorbent 
[illegible] and sensory [illegible].
f. Place apparatus and capacity switching unit in cartoon packages 
and designated bags, and remove air by pressing on bags in 
package. After that seal bag from top, the edge should be strait 
with no curves.
g. Top portion of bag maybe welded by designated electric iron; 
for purposes of easy operation, iron front head is bent with an 
angle of 50-60 degrees for a distance of 25-30 millimeter from end 
of iron head. Unite iron end with [illegible] to achieve a smooth 
surface with a width of 5 millimeter and length 25-30 millimeter. 
Welded id done as follows:
a. Place packaging box on a smooth table covered with several 
layers of thick paper.
b. Map inside surface of bag and at place of welding by a rag 
wetted with gas (be careful of fire).
c. Place paper on outside edges of bag and weld edges with a 
distance of 5 millimeters of final edges of paper bag (a regular 
iron maybe used for welding).
Rear holes of apparatus are closed when transported and not in the 
wheel tied to.

	213
	RESTRICTED

	RESTRICTED

Supplement 2

Components to be replaced after they expire

SERIAL COMPONENT NAME SYMBOL OPERATION TIME (HOUR) NOTES 
1.Electro-light adapter light KTM-620002.Air measurement light 
CT-24-1,220003.Green light on control board and light signals 
CT-24-1,220004.Blue light on control board and light signals and 
mobile signals CT-24-1,22000
NOTE: Components which are due for replacement after they expire 
are taken from reserve material unit (illegible) during 
maintenance and based upon what is written in service book.

	214
	RESTRICTED

	RESTRICTED

Supplement 3
Naming and designating components used during maintenance

SERIAL WORK NAME NAME OF MATERIAL USED DURING MAINTENANCE 
ALLOCATING AMOUNT OF MATERIAL USED 1.Checking general view Gas, 
rag 0.2 liter 2. Checking technical condition of apparatus Sensory 
tape
Distilled water 5 meters
0.3 liter 3.Washing gas passageway of apparatus Alcohol 0.3 liter 
4. Storage and lubrication Gas
Rag
Oil 0.2 liter
0.2 klg.
0.02 klg.


	215
	RESTRICTED

	RESTRICTED



1. Connecting the sockets is indicated in a symbolic way on both 
the wire and the socket, example (1) is connected with socket 
(LLL-1) and so on.
2. The voltage is indicated on the device.
3. The side network is allowed to be connected with socket LLL-2 
to supply the supporting devices in case the device structure is 
not connected with the network.
4. Choose resistance (1 - R sub 5) when regulating. 









































	figure number 2
Keys: 
1. the relay block
2. device front
3. battery box
4. transferred signals block
5. circuit 



	RESTRICTED